Chapter 16 559
Specifications and Supplemental Information
Source control
16. Specifications and
Supplemental Information
Source control
Measured number of points per sweep User definable from 2 to 1601.
Sweep mode Normal stepped, normal swept, fast stepped and fast swept.
Sweep type Linear sweep, segment sweep, log sweep and power sweep.
Segment sweep Define independent sweep segments. Set number of points, test port
power levels, IF bandwidth, delay time, sweep time and sweep mode
independently for each segment.
Sweep trigger Set to continuous, hold, or single, sweep with internal, external, manual,
or bus trigger.
Power Set source power from -15 dBm (-50 dBm for option 214/314/414) to 10
dBm. The power slope function and the power calibration function
compensate source power level error.
Trace functions
Display data Display current measurement data, memory data, or current
measurement and memory data simultaneously.
Trace math Vector addition, subtraction, multiplication or division of measured
complex values and memory data.
Title Add custom title to each channel window. Titles are printed on
hardcopies of displayed measurements.
Autoscale Automatically selects scale resolution and reference value to vertically
center the trace.
Electrical delay Offset measured phase or group delay by a defined amount of electrical
delay, in seconds.
Phase Offset Offset measured phase or group delay by a defined amount in degrees.
Statistics Calculates and displays mean, standard deviation and peak-to-peak
deviation of the data trace.