EasyManua.ls Logo

Agilent Technologies ENA Series

Agilent Technologies ENA Series
769 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 4 193
Calibration
Changing the Calibration Kit Definition
4. Calibration
Default settings of pre-defined calibration kits
The calibration kits 85033E, 85033D, 85052D, 85032F, 85032B, 85036B/E, 85031B,
85050C/D, and 85052C are pre-defined with their own default settings.
85033E
1. Short 2. Open 3. Broadband 4. Thru
Label Short Open Broadband Thru
STD Type Short Open Load Delay/Thru
C0 [×10
-15
F]
0 49.43 0 0
C1 [×10
-27
F/Hz]
0 -310.13 0 0
C2 [×10
-36
F/Hz
2
]
0 23.17 0 0
C3 [×10
-45
F/Hz
3
]
0 -0.16 0 0
L0 [×10
-12
H]
2.0765 0 0 0
L1 [×10
-24
H/Hz]
-108.54 0 0 0
L2 [×10
-33
H/Hz
2
]
2.1705 0 0 0
L3 [×10
-42
H/Hz
3
]
0.01 0 0 0
Offset Delay [s] 31.808 p 29.243 p 0 0
Offset Z0 [Ω] 50 50 50 50
Offset Loss [Ω/s] 2.36 G 2.2 G 2.2 G 2.2 G
Arb. Impedance [Ω] 50 50 50 50
85033D
1. Short 2. Open 3. Broadband 4. Thru
Label Short Open Broadband Thru
STD Type Short Open Load Delay/Thru
C0 [×10
-15
F]
0 49.43 0 0
C1 [×10
-27
F/Hz]
0 -310.13 0 0
C2 [×10
-36
F/Hz
2
]
0 23.17 0 0
C3 [×10
-45
F/Hz
3
]
0 -0.16 0 0
L0 [×10
-12
H]
2.0765 0 0 0
L1 [×10
-24
H/Hz]
-108.54 0 0 0
L2 [×10
-33
H/Hz
2
]
2.1705 0 0 0
L3 [×10
-42
H/Hz
3
]
0.01 0 0 0
Offset Delay [s] 31.808 p 29.243 p 0 0
Offset Z0 [Ω] 50 50 50 50
Offset Loss [Ω/s] 2.36 G 2.2 G 2.2 G 2.2 G
Arb. Impedance [Ω] 50 50 50 50

Table of Contents

Other manuals for Agilent Technologies ENA Series

Related product manuals