Chapter 4 189
Calibration
Changing the Calibration Kit Definition
4. Calibration
Step 7. In the same way, repeat Step 2 to Step 6 to enter the definition of REFLECT to No. 2.Select
SHORT for STD Type.
Procedure to define Match
Step 1. Press 3:No Name - Label.
Step 2. Type in MATCH <2G.
Step 3. Press STD Type - Load.
Step 4. Set Max Frequency to 2GHz.
Step 5. Press Return to return to the Define Std menu.
Procedure to define Line 1/2
Step 1. Press 4:No Name - Label.
Step 2. Type in LINE <7G.
Step 3. Press STD Type - Thru.
Step 4. Set Offset Delay to 54ps.
Step 5. Set Min Frequency to 2GHz.
Step 6. Set Max Frequency to 7GHz.
Step 7. Press Return to return to the Define Std menu.
Step 8. Press 5:No Name - Label.
Step 9. Type in LINE >7G.
Step 10. Press STD Type - Thru.
Step 11. Set Offset Delay to 13ps.
Step 12. Set Min Frequency to 7GHz.
Step 13. Press Return to return to the Define Std menu.
Step 14. Press Return to return to the Modify Cal Kit menu.
Procedure to register Thru/Reflect/Match to sub class 1
Step 1. Press Specify CLSs - Sub Class.
Step 2. Select Sub Class1.
Step 3. Press TRL Thru - Set All - THRU - Return.
Step 4. Press TRL Reflect - REFLECT.
Step 5. Press TRL Line/Match - Set All - MATCH <2G - Return.
Procedure to register Line 1/2 to subclass 2/3
Step 1. Select Sub Class2.
Step 2. Press TRL Line/Match - Set All - LINE <7G - Return.