Chapter 4 199
Calibration
Changing the Calibration Kit Definition
4. Calibration
85050C/D
1. Short(m) 2. Open(m) 3. Broadband 4. Thru
Label Short Open Broadband Thru
STD Type Short Open Load Delay/Thru
C0 [×10
-15
F]
0 90.48 m 0 0
C1 [×10
-27
F/Hz]
0 763.3 m 0 0
C2 [×10
-36
F/Hz
2
]
0 -63.82 m 0 0
C3 [×10
-45
F/Hz
3
]
0 6.434 m 0 0
L0 [×10
-12
H]
356.6 m 0 0 0
L1 [×10
-24
H/Hz]
-33.39 0 0 0
L2 [×10
-33
H/Hz
2
]
1.754 0 0 0
L3 [×10
-42
H/Hz
3
]
-33.60 m 0 0 0
Offset Delay [s] 0 0 0 0
Offset Z0 [Ω] 50 50 50 50
Offset Loss [Ω/s] 0 0 0 0
Arb. Impedance [Ω] 50 50 75 50
5. 2-18 Line 6. Lowband
Label Thru Lowband
STD Type Delay/Thru Load
C0 [×10
-15
F]
0 0
C1 [×10
-27
F/Hz]
0 0
C2 [×10
-36
F/Hz
2
]
0 0
C3 [×10
-45
F/Hz
3
]
0 0
L0 [×10
-12
H]
0 0
L1 [×10
-24
H/Hz]
0 0
L2 [×10
-33
H/Hz
2
]
0 0
L3 [×10
-42
H/Hz
3
]
0 0
Offset Delay [s] 23.190 p 0
Offset Z0 [Ω] 49.988 50
Offset Loss [Ω/s] 0.7 G 0
Arb. Impedance [Ω] 50 50
NOTE The designation (m) in the name (label) of the standard indicates male analyzer connectors.