Chapter 4 201
Calibration
Changing the Calibration Kit Definition
4. Calibration
85038A/F/M
1. Short 2. Open 3. Broadband 4. Thru
Label Short Open Broadband Thru
STD Type Short Open Load Delay/Thru
C0 [×10
-15
F]
0 32.0 0 0
C1 [×10
-27
F/Hz]
0 100 0 0
C2 [×10
-36
F/Hz
2
]
0 -50.0 0 0
C3 [×10
-45
F/Hz
3
]
0 100 0 0
L0 [×10
-12
H]
0 0 0 0
L1 [×10
-24
H/Hz]
0 0 0 0
L2 [×10
-33
H/Hz
2
]
0 0 0 0
L3 [×10
-42
H/Hz
3
]
0 0 0 0
Offset Delay [s] 66.734 p 66.734 p 0 0
Offset Z0 [Ω] 50 50 50 50
Offset Loss [Ω/s] 0.63 G 0.63 G 0 0
Arb. Impedance [Ω] 50 50 50 50