IPMEDPRO (IP Media Processor Circuit Pack)
Issue 1 May 2002
8-931555-233-143
Notes:
a. Error Type 1—indicates that the circuit pack has been removed from the
system or is not fully administered. The alarm is logged about 15 minutes
after the circuit pack has been removed, or 11 minutes after the
SAKI test
(#53) fails.
To be fully administered, a IPMEDPRO circuit pack must meet all of these
four conditions:
■ Have an entry in the circuit pack form (change circuit pack)
■ Have the MedPro IP address administered (change node-names)
■ Be enabled (change ip-interface)
■ Be physically inserted into the correct slot
If the circuit pack has an entry in the circuit pack form and either of the
other two conditions are not met, a MINOR alarm is logged. To resolve the
error, either:
■ Make sure all conditions for administration are met and that a
functioning MedPro circuit pack is inserted in the correct slot.
■ Completely remove the IPMEDPRO from the system using the
following steps:
a. Remove the administered IP-Interface associated with the
circuit pack.
b. Physically remove the circuit pack from the slot.
c. Execute the remove medpro UUCSS and change circuit
pack UUCSS commands.
b. Error Type 18—The IPMEDPRO interface circuit pack has been busied
out by a busyout board UUCSS command.
1. Release the circuit pack (release board UUCSS).
1. Release the circuit pack (release board UUCSS).
c. Error Type 23—The IPMEDPRO circuit pack is not completely
administered. To be fully administered, an IPMEDPRO circuit pack must
meet all of these four conditions:
■ Have an entry in the circuit plan (change circuit pack)
■ Have the IPMEDPRO IP address administered (change
node-names)
■ Be enabled (change ip-interface)
■ Be physically inserted into the correct slot.
A DS1 (IPMEDPRO, MEDPRO, MAPD, UDS1-BD and DS1-BD) differs
from most circuit packs in that inserting the circuit pack into the switch is
not enough to make the board usable. It must also be administered.