AS Autosampler Operator’s Manual
108 Doc. 065051-03 1/08
[367] The computed sample injection volume exceeded the vial
volume. The sample injection volume has been automatically
reduced. Please confirm injection volume parameter.
Actions: See Section 3.6.4
for the guidelines for sample volumes
used, loop sizes, and injection volumes. See Section 2.3.5
for a
list of the available sample vial and well plate sizes.
[368] The computed sample injection volume exceeded the syringe
volume. The sample injection volume has been automatically
reduced. Please confirm injection volume parameter.
Actions: See Section 3.6.4
for the guidelines for sample volumes
used, loop sizes, syringe sizes, and injection volumes.
[375] An attempt was made to use the sample prep syringe. This
option is not installed.
Two of the method sample prep operations
(DISPENSE and DILUTE)
can be used only when the sample preparation option is installed. See
Section D.5.3
for details about defining sample prep steps from
the front panel.
Possible Causes: A
DISPENSE or DILUTE operation was specified
in the sample prep steps of the current method, although the
sample preparation option is not installed.
Actions:
Remove the step(s) requiring the sample preparation option.
Select a method that does not require the sample preparation
option.
[376] Time-out occurred waiting for syringe to complete an action.
Every syringe action is assigned a maximum time in which the action
should be completed. The syringe did not complete its task in the
allotted time, indicating either a mechanical or a communication
problem.
Possible Causes:
Faulty syringe