5•Service
Doc. 065051-03 1/08 159
Notes for calibrating the inject port volume when in sequential mode:
Preparing: After disconnecting the diverter valve lines from the injection
valves, route the lines outside the autosampler or chromatography
compartment.
Make sure the sample syringe is free of bubbles (see Section B.3.14
).
Step 1: Select Diverter Valve Position_1 to calibrate the inject port
volume for system 1.
Figure 5-28. Chromeleon Inject Port Volume Calibration Panel: Sequential Mode