D • Front Panel Control
Doc. 065051-03 1/08 315
29. Select the following options:
DISPENSE: 50 μL
SYRINGE SPEED: 5
ACTION
: DISPENSE
30. Press Enter.
31. Select
VIAL#: INJ and press Enter.
The needle arm moves to the inject port.
32. Select the following options:
DISPENSE: volume (microliters) recorded in Step 21
SYRINGE SPEED: 3
ACTION
: DISPENSE
33. Press Enter.
34. Verify that a drop of liquid appears at the end of the needle seal tubing. If no
drop appears or if the drop is a different size than previously observed, repeat
Step 7
through Step 34.
For sequential mode, switch the diverter valve to the other position and repeat
Step 22
through Step 33.
35. Select
VIAL#: FLU and press Enter.
36. Select the following options:
SYRINGE SPEED: 5
ACTION
: EMPTY
37. Press Enter.
38. Press
Home to move the needle arm to the home position.
39. Press
Menu twice to go to the MENU of SCREENS.