IRECTION FR091521, REVISION 1 VIVID S60N/VIVID S70N BASIC SERVICE MANUAL
8-86 Section 8-5 - Electronic Cage Components - Replacement Procedures
PRELIMINARY
6.) Release the cage supporting arm from the device chassis and return it to its location on the cage -
see Figure 8-109.
7.) Open the safety-catch on the cage-hinge - see Figure 8-110.
8.) Carefully lift the cage upwards to free it from the securing hinge (see Figure 8-111) and place it on
a clean, stable surface.
Figure 8-109 Releasing Cage Supporting Arm from Device Chassis
Figure 8-110 Open Safety-Catch on Hinge of Cage Door
Figure 8-111 Removing the Front End Door Assembly from the Securing Hinge