After touching to disable the mem-
ory function, use or to set the num-
ber of measurement results.
Settable range: 1 to 32000
The number of measurement results can only be set
when the memory function is set to OFF.
3
Memory Function Setting
Select ON/OFF for the memory function.
Disables the memory function.
Saves all measurement values to memory.
Clears all of the saved measurement values
from the instrument memory.
Saves the measurement values stored in the in-
strument memory to a USB flash drive and then
clears the measurement values from the instru-
ment memory.
The measurement values are saved to the
MEMORY folder in the USB flash drive. The file
name is automatically assigned from the date
and time (p. 285).
4
Memory Function Setting
5
Press to close the setting screen.