4.2 Setting Basic Settings of Measurement Conditions
79
4
Chapter 4 LCR Function
The trigger delay function allows you to set a delay time from the time the trigger signal is input until measure-
ment. The trigger synchronous output function outputs the measurement signal only during measurement and
allows you to set a delay time that will be allowed to elapse before data is acquired.
The measurement process is as follows:
Trigger delays and the trigger synchronous output function
Measurement
Data acquisition
Trigger Delay
Measurement
signal
Trigger Signal
Trigger Signal
Measurement
signal
Trigger delay: on; trigger synchronous output: on
Trigger Delay
Trigger synchro-
nization delay
Data acquisition
Trigger delay: on; trigger synchronous output: off
Measurement
ON
ON
When the range synchronization function has been set, the range settings at which the trigger
delay and trigger synchronous output function are enabled vary with the parameter settings.
Parameter
Range setting at which
function is enabled
AC measurement only AC measurement range
AC+DC measurement AC measurement range
DC measurement only DC measurement range