EasyManua.ls Logo

Keysight U8480 Series - Page 82

Keysight U8480 Series
375 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
2 MEASurement Commands
82 U8480 Series Programming Guide
Parameters
Refer to “Optional parameters” on page 73 for additional details on the
parameters in this command.
Error messages
If this query is sent, and TRIG:COUNT > 1, error –221, “Settings conflict”
occurs.
Item Description/Default Range of values
expected_value The expected power level parameter can be set to DEF or a
numeric value. If a value is entered, it should correspond to
that set by CONFigure otherwise an error occurs. The units of
measurement are dBm and W. The default units are defined by
UNIT:POWer.
sensor-dependent
DEF
[a]
resolution A numeric value for the resolution. If it is unspecified, the
current resolution setting is used. If a value is entered, it
should correspond to the current resolution setting otherwise
an error occurs.
1 to 4
[b]
1.0, 0.1, 0.01, 0.001
DEF
[a]
source list The measurement channel which the command is
implemented on. The U8480 Series supports only one
channel. Therefore, the only valid value is (@1).
(@1)
[a] The mnemonic DEF means DEFault. This is not equivalent to the DEFault parameter used in the command subsystems. The parameters must
be entered in the specified order. If parameters are omitted, they default from the right. The parameter DEFault is used as a placeholder.
Specifying DEF leaves the parameter value unchanged.
[b] When the measurement result is linear, this parameter represents the number of significant digits. When the measurement result is
logarithmic, 1 to 4 represents 1, 0.1, 0.01, and 0.001 respectively.

Table of Contents

Related product manuals