Adjusting the camera (only with VisionSystem)
134 ProtoMat
®
S62
6.5.1 Calibrating fiducial recognition
Fiducial recognition is factory-set for bores 1.5 mm in diameter. The set-
tings must be recalibrated if the fiducials are not clearly recognised.
I.
Preparing the fiducial recognition
1. Drill a hole into an unused area of the material to be used with the
Spiral Drill 1.5 mm.
2. Click on Move to > Camera>>Head in the main menu.
The camera is positioned above the bore and the bore is visible in the
Camera image window.
3. Click on Setting.
The Setting window is opened.
Fig. 71: Bore in the camera
image window