Remote Control
R&S
®
ESR
1036User Manual 1175.7068.02 ─ 12
Result
Figure 11-12: Result: noise only
Measurement example: Periodic signal intervals in edge mode
The gate area can also be defined by a number of gate periods and the distance
between two periods (gap). In this example, a power ramp is recorded, triggered by the
IFP on the highest edge (see Figure 11-13). As a result, only the levels of the individual
ramp areas of the first 3 periods are displayed (see Figure 11-14).
Figure 11-13: Periodic signal recording in edge mode
Used values:
sample rate: 3,2 MHz
number of IQ samples: 228000
trigger mode: IFP
Remote Control – Programming Examples

Table of Contents