Receiver Mode
R&S
®
ESR
161User Manual 1175.7068.02 ─ 12
Writes the maximum level that has been measured on this frequency during the
prescan to the final peak list without performing a final measurement.
● Stop the final measurement ("Stop Final Meas")
4. The level measurement on the current frequency is initiated after you press the
"Measure" softkey.
5. After the final measurement on the current frequency is done, the R&S ESR repla-
ces the scan result in the peak list with the result of the final measurement. If the
frequency has drifted compared to the one of the prescan, it also updates the fre-
quency in the peak list.
6. The R&S ESR moves to the next frequency in the peak list, positions the marker
on that frequency etc.
7. After all frequencies in the peak list are finished, the R&S ESR opens the "Final
Peak List" with the results for the final measurement.
Note that it is possible to start with an automatic measurement and later change into
interactive mode. Likewise, it is possible to start measuring in interactive mode and
later change into automatic mode.
Hold Final Measurement.............................................................................................161
Automatic Final........................................................................................................... 161
Interactive Final...........................................................................................................161
Skip Frequency........................................................................................................... 162
Get Maxhold................................................................................................................162
Measure...................................................................................................................... 162
Stop Final Measurement.............................................................................................162
Hold Final Measurement
Interrupts the final measurement and opens a submenu that contains functionality to
control the final measurement.
Data that has already been collected is kept in the memory.
Remote command:
HOLD on page 640
Automatic Final
Selects an automatic final measurement.
See also Chapter 5.1.5, "Final Measurement", on page 150.
Remote command:
[SENSe:]FMEasurement:AUTO on page 643
Interactive Final
Selects interactive final measurements.
See also Chapter 5.1.5, "Final Measurement", on page 150.
Remote command:
[SENSe:]FMEasurement:AUTO on page 643
Measurements and Result Displays