Remote Control
R&S
®
ESR
645User Manual 1175.7068.02 ─ 12
4 byte, bit 0 represents the state of trace 4 (0/1)
Note:
If more than 4 traces are active, the state of trace 5 and 6 is indicated by an addi-
tional bit (bit 8) in the UINT32 field of trace 1 and 2.
n*4 byte, measurement results for trace 1; only if trace 1 is active
n*4 byte, measurement results for trace 2; only if trace 2 is active
n*4 byte, measurement results for trace 3; only if trace 3 is active
n*4 byte, measurement results for trace 4; only if trace 4 is active
n*4 byte, measurement results for trace 5; only if trace 5 is active
n*4 byte, measurement results for trace 6; only if trace 6 is active
n*1 byte, status information for each measurement result
bit 2 represents overrange for trace 1 to trace 6
The data is always returned in binary format (FORM REAL,32).
Note that the SCAN parameter only works while the scan is actually running.
Example:
TRAC? SCAN
Querying results for a peak search
The SCPI parameters PLIST1 | ... | PLIST6 returns the results of a peak search for a
particular trace or detector (1 to 6).
Each result is made up out of three values:
4 byte, frequency
4 byte, absolute level
4 byte, Delta between absolute level and limit line value at corresponding fre-
quency
If no limit line is active, the delta value is set to 0.0
The trace has to be active for the command to work.
Example:
TRAC? PLIST3
Querying results for the final measurement
The SCPI parameters FINAL1 | ... | FINAL6 return the results of the final measure-
ment for a particular trace or detector (1 to 6).
Each result is made up out of three values:
4 byte, frequency
4 byte, absolute level
4 byte, Delta between absolute level and limit line value at corresponding fre-
quency
If no limit line is active, the delta value is set to 0.0
The trace has to be active for the command to work.
Remote Commands in Receiver Mode

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S ESR Series and is the answer not in the manual?

R&S ESR Series Specifications

General IconGeneral
BrandR&S
ModelESR Series
CategoryReceiver
LanguageEnglish

Related product manuals