Remote Control
R&S
®
ESR
677User Manual 1175.7068.02 ─ 12
Parameters:
<Ranges> Number of ranges in the scan table.
If you enter the value "0", the R&S ESR ignores the configura-
tion of the scan table. instead, it performs the measurement
based on the current receiver configuration.
Range: 0 to 10
*RST: 0
Example:
SCAN:RANG:COUN 4
Defines 4 scan ranges.
[SENSe:]SCAN<range>:STARt <Frequency>
This command defines the start frequency of a particular scan range.
Suffix:
<range>
.
1...10
Selects the scan range.
Parameters:
<Frequency> Range: fmin to fmax
*RST: 150 kHz
Example:
SCAN4:STAR 30MHz
Defines a start frequency of 30 MHz for the 4th scan range.
[SENSe:]SCAN<range>:STEP <Frequency>
This command defines the frequency stepsize within a particular scan range.
Suffix:
<range>
.
1...10
Selects the scan range.
Parameters:
<Frequency> Range: fmin to fmax
*RST: 4 kHz
Example:
SCAN4:STEP 1MHz
Defines a step size of 1 MHz for the 4th scan range.
[SENSe:]SCAN<range>:STOP <Frequency>
This command defines the stop frequency of a particular scan range.
Suffix:
<range>
.
1...10
Selects the scan range.
Parameters:
<Frequency> Range: fmin to fmax
*RST: 30 MHz
Remote Commands in Receiver Mode

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S ESR Series and is the answer not in the manual?

R&S ESR Series Specifications

General IconGeneral
BrandR&S
ModelESR Series
CategoryReceiver
LanguageEnglish

Related product manuals