[SENSe:]SCAN<range>:INPut:GAIN:AUTO <State>
This command includes the preamplifier in the autoranging function of the selected
receiver scan range.
Parameters:
<State> ON
OFF
*RST: OFF
Example:
:SCAN1:INP:GAIN:AUTO ON
Includes the preamplifier in the autoranging function for scan
range 1
[SENSe:]SCAN<range>:INPut:GAIN[:STATE]
<State>
This command turns the preamplifier in a particular scan range on and off.
Suffix:
<range>
.
1...10
Selects the scan range.
Parameters:
<State> ON | OFF
*RST: OFF
Example:
SCAN4:INP:GAIN:STAT ON
Turns on the preamplifier for the 4th scan range.
[SENSe:]SCAN<range>:INPut:TYPE <Input>
This command selects the signal source for a particular scan range.
Suffix:
<range>
.
1...10
Selects the scan range.
Parameters:
<Input> INPUT1
Selects RF input 1.
INPUT2
Selects RF input 2.
*RST: INPUT1
Example:
SCAN4:INP:TYPE INPUT2
Selects RF input 2 as the signal source for the 4th scan range.
[SENSe:]SCAN:RANGes[:COUNt] <Ranges>
This command defines the number of scan ranges.
Remote Commands in Receiver Mode