EasyManua.ls Logo

R&S ESR Series

R&S ESR Series
1126 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control
R&S
®
ESR
676User Manual 1175.7068.02 ─ 12
[SENSe:]SCAN<range>:INPut:GAIN:AUTO <State>
This command includes the preamplifier in the autoranging function of the selected
receiver scan range.
Parameters:
<State> ON
OFF
*RST: OFF
Example:
:SCAN1:INP:GAIN:AUTO ON
Includes the preamplifier in the autoranging function for scan
range 1
[SENSe:]SCAN<range>:INPut:GAIN[:STATE]
<State>
This command turns the preamplifier in a particular scan range on and off.
Suffix:
<range>
.
1...10
Selects the scan range.
Parameters:
<State> ON | OFF
*RST: OFF
Example:
SCAN4:INP:GAIN:STAT ON
Turns on the preamplifier for the 4th scan range.
[SENSe:]SCAN<range>:INPut:TYPE <Input>
This command selects the signal source for a particular scan range.
Suffix:
<range>
.
1...10
Selects the scan range.
Parameters:
<Input> INPUT1
Selects RF input 1.
INPUT2
Selects RF input 2.
*RST: INPUT1
Example:
SCAN4:INP:TYPE INPUT2
Selects RF input 2 as the signal source for the 4th scan range.
[SENSe:]SCAN:RANGes[:COUNt] <Ranges>
This command defines the number of scan ranges.
Remote Commands in Receiver Mode

Table of Contents

Related product manuals