Remote Control
R&S
®
ESR
718User Manual 1175.7068.02 ─ 12
Suffix:
<n>
.
irrelevant
<k> irrelevant
Example:
CALC:LIM:CLE
Deletes the result of the limit check.
Usage: Event
CALCulate<n>:LIMit<k>:FAIL?
This command queries the result of a limit check.
Note that for SEM measurements, the limit line suffix <k> is irrelevant, as only one spe-
cific SEM limit line is checked for the currently relevant power class.
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweeps.
Suffix:
<n>
.
irrelevant
<k> limit line
Return values:
<Result> 0
PASS
1
FAIL
Example:
INIT;*WAI
Starts a new sweep and waits for its end.
CALC:LIM3:FAIL?
Queries the result of the check for limit line 3.
Usage: Query only
Manual operation: See "Limit Check 1-4" on page 261
See "Limit Check 1-4" on page 291
CALCulate<n>:LIMit<k>:STATe <State>
This command turns the limit check on and off.
To query the limit check result, use CALCulate<n>:LIMit<k>:FAIL?.
Suffix:
<n>
.
irrelevant
<k> Selects the limit line.
Parameters:
<State> ON | OFF
*RST: OFF
Remote Commands in Receiver Mode

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S ESR Series and is the answer not in the manual?

R&S ESR Series Specifications

General IconGeneral
BrandR&S
ModelESR Series
CategoryReceiver
LanguageEnglish

Related product manuals