CALCulate<n>:LIMit<k>:ESPectrum:RESTore
This command restores the predefined limit lines for the Spectrum Emission Mask
measurement. All modifications made to the predefined limit lines are lost and the fac-
tory-set values are restored.
Suffix:
<n>
.
1...4
window
<k> irrelevant
Example:
CALC:LIM:ESP:REST
Resets the limit lines for the Spectrum Emission Mask to the
default setting.
[SENSe:]ESPectrum:PRESet[:STANdard]
This command selects the specified XML file under C:\r_s\instr\sem_std. If the
file is stored in a subdirectory, include the relative path.
Example:
ESP:PRES 'WCDMA\3GPP\DL\PowerClass_31_39.xml'
Selects the PowerClass_31_39.xml XML file in the C:
\R_S\instr\sem_std\WCDMA\3GPP\DL directory.
ESP:PRES?
W-CDMA 3GPP DL (31,39)dBm
The query returns information about the selected standard, the
link direction and the power class. If no standard has been
selected, the query returns None.
Manual operation: See "Load Standard" on page 267
[SENSe:]ESPectrum:PRESet:RESTore
This command copies the XML files from the C:\R_S\instr\sem_backup folder to
the C:\R_S\instr\sem_std folder. Files of the same name are overwritten.
Example:
ESP:PRES:REST
Restores the originally provided XML files.
Manual operation: See "Restore Standard Files" on page 267
[SENSe:]ESPectrum:PRESet:STORe <FileName>
This command stores the current settings as presettings in the specified XML file under
C:\r_s\instr\sem_backup.