Remote Control
R&S
®
ESR
773User Manual 1175.7068.02 ─ 12
CALCulate<n>:LIMit<k>:ESPectrum:PCLass<Class>:MAXimum........................................ 775
CALCulate<n>:LIMit<k>:ESPectrum:PCLass<Class>:MINimum..........................................775
CALCulate<n>:LIMit<k>:ESPectrum:VALue...................................................................... 776
CALCulate<n>:LIMit<k>:ESPectrum:LIMits <Limits>
This command sets or queries up to 4 power classes in one step.
Suffix:
<n>
.
irrelevant
<k> irrelevant
Parameters:
<Limits> 1–3 numeric values between -200 and 200, separated by com-
mas
-200, <0-3 numeric values between -200 and 200, in ascending
order, separated by commas>, 200
Example:
CALC:LIM:ESP:LIM -50,50,70
Defines the following power classes:
<-200, -50>
<-50, 50>
<50, 70>
<70, 200>
Query:
CALC:LIM:ESP:LIM?
Response:
-200,-50,50,70,200
CALCulate<n>:LIMit<k>:ESPectrum:MODE <Mode>
This command activates or deactivates the automatic selection of the limit line in the
Spectrum Emission Mask measurement.
Suffix:
<n>
.
1...4
window
<k> irrelevant
Parameters:
<Mode> AUTO | MANUAL
AUTO
The limit line depends on the measured channel power.
MANUAL
One of the three specified limit lines is set.
*RST: AUTO
Example:
CALC:LIM:ESP:MODE AUTO
Activates automatic selection of the limit line.
Remote Commands in Spectrum Analyzer Mode

Table of Contents