Remote Control
R&S
®
ESR
816User Manual 1175.7068.02 ─ 12
Instrument setup and querying of the result list at the end of the measurement:
With this method, the control computer may be used for other activities while the
measurement is being performed. However, more time is needed for synchroniza-
tion via service request.
[SENSe:]LIST:POWer:RESult?........................................................................................ 816
[SENSe:]LIST:POWer[:SEQuence]...................................................................................817
[SENSe:]LIST:POWer:SET..............................................................................................819
[SENSe:]LIST:POWer:STATe...........................................................................................820
[SENSe:]LIST:POWer:RESult?
This command queries the results of the list evaluation.
This command may be used to obtain measurement results in an asynchronous way,
using the service request mechanism for synchronization to the end of the measure-
ment.
If there are no results, the command returns an error.
Return values:
<PowerLevel> Power level for each frequency included in the measurement.
The command returns up to 3 power levels for each frequency,
depending on the number of evaluation modes you have turned
on with [SENSe:]LIST:POWer:SET on page 819.
The result is a list of floating point values separated by commas.
The unit depends on [SENSe:]LIST:POWer[:SEQuence].
Example:
*ESE 1
*SRE 32
Configuration of the status reporting system for the generation of
an SRQ on operation complete
SENSe:LIST:POWer
935.2 MHz,-20dBm,10dB,OFF,NORM,1 MHz,3 MHz,
434us,0,
935.4 MHz,-20dBm,10dB,10dB,NORM,30 kHz,100 kHz,
434us,0,
935.6 MHz,-20dBm,10dB,20dB,NORM,30 kHz,100 kHz,
434us,0;
*OPC
Configuring and starting the measurement
...
Further actions of the control computer during measurement
On SRQ:
SENSe:LIST:POWer:RESult?
Response to service request
Usage:
Query only
Remote Commands in Spectrum Analyzer Mode