Remote Control
R&S
®
ESR
857User Manual 1175.7068.02 ─ 12
Example:
CALC:SGR:COL GRAY
Selects black and white color scheme.
Manual operation: See "Color Mapping" on page 406
CALCulate<n>:SGRam:CONT <State>
This command determines whether the results of the last measurement are deleted
before starting a new measurement in single sweep mode.
Suffix:
<n>
.
1...4
Selects the measurement window.
Parameters:
<State> ON | OFF
*RST: OFF
Example:
INIT:CONT OFF
Selects single sweep mode.
INIT;*WAI
Starts the sweep and waits for the end of the sweep.
CALC:SGR:CONT ON
Repeats the single sweep measurement without deleting the
results of the last measurement.
Manual operation: See "Continue Frame (On Off)" on page 373
CALCulate<n>:SGRam:FRAMe:COUNt <Frames>
This command sets the number of frames to be recorded in a single sweep.
Suffix:
<n>
.
1...4
Selects the measurement window.
Parameters:
<Frames> The maximum number of frames depends on
the .CALCulate<n>:SGRam:HDEPth on page 651
Range: 1 to depends on history depth
Increment: 1
*RST: 1
Example:
INIT:CONT OFF
Selects single sweep mode.
CALC:SGR:FRAM:COUN 200
Sets the number of frames to 200.
Manual operation: See "Frame Count" on page 373
Remote Commands in Spectrum Analyzer Mode