EasyManua.ls Logo

Ricoh A66 - Touch Panel Calibration

Ricoh A66
733 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
12.4 TOUCH PANEL CALIBRATION
A189 Copier Only
After replacing the operation panel, the touch panel must be calibrated as
follows:
1. Enter SP5-901 "Touch Panel Calibration" and select "0" (reset). This step
cancels the isting touch panel calibration and allows the panel to be
re-calibrated.
2. The operation panel displays four screens in sequence, as shown.
Gently touch the center of the "X" on each screen with the tip of a pen.
NOTE: Do not use too sharp a tip in this step. The touch panel may be
damaged.
3. The operation panel displays the final screen shown. Point at any place
on this screen and make sure that a "O" mark appears at that point. If the
mark appears properly, press [Go].
4. it the SP mode.
Cancel Go
1.
2.
3.
4.
Replacement
and
Adjustment
OTHERS
SM 5-101 A166/A187/A189

Table of Contents