◼ Tall T-wave rejection capability: 0.2~1.2mV
◼ Pacemaker pulse rejection capability: ±2~700mV, 0.1~2.0ms
◼ - FILTER -
1) 0.5Hz selection: used when baseline movement is severe
and when ECG diagnosis (arrhythmia, ST) is not needed
2) 0.05 Hz selection: when baseline movement is not severe
and when an electrocardiogram (arrhythmia, ST) is needed
3) 40Hz selection: mainly used for human body
4) 80 Hz selection: for animals (the frequency range of the QRS
complex is wider than the human body).
5) Monitoring: 0.5 ~ 40Hz filter + 5- 6th moving average filter
(in case of very noisy environment)
◼ Do not use disposable electrode repeatedly. (Skin trouble, They
may cause noise.)
◼ Wash patient’s skin before using electrode. When patient put
lotion, oil & etc, adhesive strength of electrode can be lower
and ECG response time can be longer.