VM15 User Manual ECG Monitoring
113
NOTE
Tall T-wave rejection capability: 0.2~1.2mV
Pacemaker pulse rejection capability: ±2~700mV, 0.1~2.0ms
- FILTER -
1) 0.5Hz selection: used when baseline movement is severe
and when ECG diagnosis (arrhythmia, ST) is not needed
2) 0.05 Hz selection: when baseline movement is not severe
and when an electrocardiogram (arrhythmia, ST) is needed
3) 40Hz selection: mainly used for human body
4) 80 Hz selection: for animals (the frequency range of the QRS
complex is wider than the human body).
5) Monitoring: 0.5 ~ 40Hz filter + 5- 6th moving average filter
(in case of very noisy environment)
Do not use disposable electrode repeatedly. (Skin trouble, They
may cause noise.)
Wash patient’s skin before using electrode. When patient put
lotion, oil & etc, adhesive strength of electrode can be lower
and ECG response time can be longer.