VM15 User Manual Masimo SpO2
Monitoring (Option)
138
monitors or with licensed Masimo SET pulse oximetry modules during no motion.
Numbers represent ± 1 standard deviation. Plus or minus one standard deviation represents
68% of the population. SpO2 accuracy from 70% to 100%. Pulse Rate accuracy from 25 to
240bpm.
CLEANING AND REUSE OF MASIMO REUSABLE SENSORS AND CABLES
Reusable sensors and patient cables can be cleaned per the following procedure:
I. Remove the sensor from the patient.
II. Disconnect the sensor from the patient cable.
III. Disconnect the patient cable from the monitor.
IV. Wipe the entire sensor and/or patient cable clean with a 70% isopropyl alcohol pad.
V. Allow to air dry thoroughly before returning it to operation.
CAUTION: CAREFULLY ROUTE PATIENT CABLES TO REDUCE THE POSSIBILITY OF
PATIENT ENTANGLEMENT OR STRANGULATION.
REATTACHMENT OF SINGLE USE ADHESIVE SENSORS
Single use sensors may be reapplied to the same patient if the emitter and detector
windows are clear and the adhesive still adheres to the skin.
If the sensor fails to track the pulse consistently, the sensors may be
incorrectly
positioned. Reposition the sensor or choose a different monitoring site.
CAUTION: D O NOT REPROCESS ANY MASIMO SINGLE USE SENSORS.
Messages
Information messages
The following chart alphabetically lists all system messages displayed on the LCD screen.
The cause of the message, and the action(s) to be taken are also shown.
The operator should become thoroughly familiar with this information before using the
oximeter for patient monitoring
Table 20-4. Information messages
Broken sensor cable wire or inoperative LEDs
or faulty detector; Replace sensor.
Refer to the instructions for the sensor being
used.