VM15 User Manual Specification
238
NIBP(Non-invasive Blood Pressure Measurement) – Standard
⚫ Measurement Method: Oscillometric, Automatic measurement
⚫ Display Value: Systolic, Diastolic, Mean Pressure, Pulse Rate
⚫ Measurement Range
◼ Adult, Pediatric: 0 to 300mmHg
◼ Neonate: 0 to 150mmHg
⚫ Accuracy
◼ Systolic, Diastolic, Mean: ± 5 mmHg
◆ Max Mean error: ± 5 mmHg
◆ Max Standard Deviation: 8 mmHg
⚫ Inflation Pressure Max
◼ Adult, Pediatric: 100 to 300mmHg
◼ Neonate: 70 to 150mmHg
◼ Venous Stasis Pressure: 50 ~ 200 mmHg
⚫ Alarm Range
NIBP
[mmHg]
Display
High
Low
Systolic
0 ~ 275, OFF
OFF, 0 ~ 275
Diastolic
Mean
⚫ Pressure Modes
◼ Manual
◼ Automatic: 2, 3, 4, 5, 10, 15, 30 min, 1, 2, 4, 8 ,12 hours intervals
◼ Stat: 5 ~ 15 min intervals
◼ Venous Stasis
⚫ Error Messages
◼ NIBP ERR. CUFF/HOSE
◼ NIBP ERR. NO PULSE
◼ NIBP ERR. PATIENT MOTION
◼ NIBP ERR. ENVELOPE
◼ NIBP ERR. WEAKPULSE
◼ NIBP ERR. TIMEOUT
◼ NIBP ERR. CUFF OVERPRESSURE
◼ NIBP ERR. INTERNAL