NIBP(Non-invasive Blood Pressure Measurement) Standard
Measurement Method: Oscillometric, Automatic measurement
Display Value: Systolic, Diastolic, Mean Pressure, Pulse Rate
Measurement Range
Adult, Pediatric: 0 to 300mmHg
Neonate: 0 to 150mmHg
Accuracy
Systolic, Diastolic, Mean: ± 5 mmHg
Max Mean error: ± 5 mmHg
Max Standard Deviation: 8 mmHg
Inflation Pressure Max
Adult, Pediatric: 100 to 300mmHg
Neonate: 70 to 150mmHg
Venous Stasis Pressure: 50 ~ 200 mmHg
Alarm Range
NIBP
[mmHg]
Display
High
Low
Systolic
0 ~ 275, OFF
OFF, 0 ~ 275
Diastolic
Mean
Pressure Modes
Manual
Automatic: 2, 3, 4, 5, 10, 15, 30 min, 1, 2, 4, 8 ,12 hours intervals
Stat: 5 ~ 15 min intervals
Venous Stasis
Error Messages
NIBP ERR. CUFF/HOSE
NIBP ERR. NO PULSE
NIBP ERR. PATIENT MOTION
NIBP ERR. ENVELOPE
NIBP ERR. WEAKPULSE
NIBP ERR. TIMEOUT
NIBP ERR. CUFF OVERPRESSURE
NIBP ERR. INTERNAL