VM15 User Manual Specification
244
TEMP - (Standard : 1 Channel / Optional : 3 Channels)
⚫ Channel: Max 4 channels (Standard: 1 channel, Optional: 3 channels)
⚫ Measurement Range: 0 ~ 50℃
⚫ Unit: ℃, ℉
⚫ Accuracy
◼ 0℃ ~ 24.9℃ → ± 0.2℃
◼ 25.0℃ ~ 50.0℃ → ± 0.1℃
⚫ Alarm Range
◼ High alarm limit: 0.1℃ ~ 50℃, OFF
◼ Low alarm limit: OFF, 0.1℃ ~ 50℃