VM15 User Manual Alarms and Limits
80
Alarm Range.
Alarm Setting Range
The Table 9-3 shows the limitation of Alarm Warning
Table 9-3. Alarm Setting Range Limit
Parameter
Setting Range
HIGH
LOW
TYPE
ECG [bpm]
0 ~ 300, OFF
OFF, 0 ~ 300
LOW,
MEDIUM,
HIGH
ST[mV]
-9.9 ~ 9.9, OFF
OFF, -9.9 ~ 9.9
VOTEM
SpO2 [%]
0 ~ 100, OFF
OFF, 0 ~ 100
SpO2 PR [bpm]
30 ~ 300, OFF
OFF, 30 ~ 300
MASIMO
SpO2 [%]
0 ~ 100, OFF
OFF, 0 ~ 100
SpO2 PR [bpm]
0 ~ 300, OFF
OFF, 0 ~ 300
VOTEM
NIBP
[mmHg]
Systolic
0 ~ 275, OFF
OFF, 0 ~ 275
Diastolic
0 ~ 275, OFF
OFF, 0 ~ 275
Mean
0 ~ 275, OFF
OFF, 0 ~ 275
SunTech
NIBP
[mmHg]
Systolic
0 ~ 275, OFF
OFF, 0 ~ 275
Diastolic
0 ~ 275, OFF
OFF, 0 ~ 275
Mean
0 ~ 275, OFF
OFF, 0 ~ 275
IBP
[mmHg]
Systolic
-50 ~ 350, OFF
OFF, -50 ~ 350
Diastolic
-50 ~ 350, OFF
OFF, -50 ~ 350
Mean
-50 ~ 350, OFF
OFF, -50 ~ 350
IBP PR [bpm]
0 ~ 300, OFF
OFF, 0 ~ 300
EtCO2 [mmHg]
0 ~ 150, OFF
OFF, 0 ~ 150
Multi-Gas
[vol%]
Gas1
In
0 ~ 25.0, OFF
OFF, 0 ~ 25.0
Ex
0 ~ 25.0, OFF
OFF, 0 ~ 25.0
Gas2
In
0 ~ 25.0, OFF
OFF, 0 ~ 25.0
Ex
0 ~ 25.0, OFF
OFF, 0 ~ 25.0
O2
In
0 ~ 100, OFF
OFF, 0 ~ 100
Ex
0 ~ 100, OFF
OFF, 0 ~ 100
N2O
In
0 ~ 100, OFF
OFF, 0 ~ 100
Ex
0 ~ 100, OFF
OFF, 0 ~ 100
CO2
In
0 ~ 25.0, OFF
OFF, 0 ~ 25.0
Ex
0 ~ 25.0, OFF
OFF, 0 ~ 25.0
qCON
0 ~ 100, OFF
OFF, 0 ~ 100.
TEMP 1 ~ 4 []
0.0 ~ 50.0, OFF
OFF, 0.0 ~ 50.0
TD
0.0 ~ 50.0, OFF
OFF, 0.0 ~ 50.0
RESP [bpm]
0 ~ 150, OFF
OFF, 0 ~ 150
RESP Apnea
OFF, 5 ~ 40
HIGH