VM15 User Manual Masimo SpO2
Monitoring (Option)
137
LNCS® ADHESIVE SENSORS
LNCS adhesive sensors must be used in conjunction with Masimo pulse oximetry.
Table 20-2. LNCS® ADHESIVE SENSORS Specification
Sensor
Weight
Range
Saturation
Accuracy
Pulse Rate
Accuracy
Low Perfusion
Accuracy
No
Motion
Motion
No
Motion
Motion
Saturation
Pulse
Rate
LNCS Adtx
LNCS Adtx-3
> 30
kg
2%
3%
3
bpm
5
bpm
2%
3
bpm
LNCS Pdtx
LNCS Pdtx-3
10 - 50
kg
2%
3%
3
bpm
5
bpm
2%
3
bpm
LNCS Inf-L
LNCS Inf
LNCS Inf-3
3 - 20
kg
2%
3%
3
bpm
5
bpm
2%
3
bpm
LNCS Neo-L
LNCS Neo
LNCS Neo-3
< 3 kg
3%
3%
3
bpm
5
bpm
3%
3
bpm
> 40
kg
2%
3%
3
bpm
5
bpm
2%
3
bpm
LNCS NeoPt-L
LNCS NeoPt
LNCS NeoPt-3
< 1 kg
3%
3%
3
bpm
5
bpm
3%
3
bpm
LNCS
NeoPt-500
< 1 kg
3%
3%
3
bpm
5
bpm
3%
3
bpm
LNCS® SPECIALTY SENSORS
LNCS specialty sensors must be used in conjunction with Masimo pulse oximetry.
Table 20-3. LNCS® SPECIALTY SENSORS Specification
Sensor
Weight
Range
Saturation
Accuracy
Pulse Rate
Accuracy
Low Perfusion
Accuracy
No
Motion
Motion
No
Motion
Motion
Saturation
Pulse
Rate
LNCS
Newborn
Infant/Pediatric
3 - 30
kg
± 2%
± 3%
± 3
bpm
± 5
bpm
± 2%
± 3
bpm
LNCS
Newborn
Neonatal
< 3 kg
± 3%
± 3%
± 5
bpm
± 5
bpm
± 3%
± 3
bpm
LNCS Trauma
> 30 kg
± 2%
± 3%
± 3
bpm
± 5
bpm
± 2%
± 3
bpm
SENSOR ACCURACY
Refer to Specifications for SpO2 and pulse rate accuracy, unless otherwise specified in the
tables above.
Accuracy specified when used with Masimo Rainbow SET technology pulse oximetry