VM15 User Manual Specification
236
SpO2 (Pulse Oximeter) - Standard
⚫ Measurement Range: 0 ~ 100 %
⚫ Accuracy
◼ 70 ~ 100 % SpO2: ± 2 %
◼ 50 ~ 69 % SpO2: ± 3 %
◼ 0 ~ 49 % SpO2: Unspecified
⚫ Pulse Rate
◼ Range: 0, 30 ~ 300 bpm
◼ Accuracy
◆ 30 ~ 240 bpm ± 2 bpm or ± 2 %, which is greater
◆ 241 ~ 300 bpm ± 3 bpm or ± 3 %, which is greater
⚫ Alarm Range
◼ High alarm limit: 0 ~ 100 % OFF
◼ Low alarm limit: OFF, 0 ~ 100 %
⚫ Averaging
◼ Set of SpO2 Measure Detect Time: 2~10 sec
⚫ Low Perfusion
◼ 0.05 % up to
⚫ Setting Time
◼ Wave out Time: Max. 2 sec
◼ SpO2 Percentage Display: Max. 10 sec