VM15 User Manual Alarms and Limits
81
Alarm Default Set
Table 9-4 below shows the default alarm settings for each patient category.
The user can adjust the alarm set value within the range as needed.
Note. The initial Patient category is set to Audlt when the product is shipped.
Table 9-4. Alarm default settings.
Parameter
Patient Category
Adult
Pediatric
Neonate
HIGH
LOW
TYPE
HIGH
LOW
TYPE
HIGH
LOW
TYPE
ECG [bpm]
120
50
HIGH
160
75
HIGH
200
100
HIGH
ST[mV]
+5.0
-5.0
MIDDLE
+5.0
-5.0
MIDDLE
+5.0
-5.0
MIDDLE
VOTEM
SpO2 [%]
OFF
90
MIDDLE
OFF
90
MIDDLE
OFF
90
MIDDLE
SpO2 PR [bpm]
120
50
HIGH
160
75
HIGH
200
100
HIGH
MASIMO
SpO2 [%]
OFF
90
MIDDLE
OFF
90
MIDDLE
OFF
90
MIDDLE
SpO2 PR [bpm]
120
50
HIGH
160
75
HIGH
200
100
HIGH
VOTEM
NIBP
[mmHg]
Systolic
160
90
HIGH
120
70
HIGH
90
40
HIGH
Diastolic
90
50
70
40
60
20
Mean
110
60
90
50
70
35
SunTech
NIBP
[mmHg]
Systolic
160
90
120
70
90
40
Diastolic
90
50
70
40
60
20
Mean
110
90
90
50
70
35
IBP 1
[mmHg]
ABP
Systolic
160
90
MIDDLE
120
70
MIDDLE
90
55
MIDDLE
Diastolic
90
50
70
40
60
20
Mean
110
60
90
50
70
35
IBP 2
[mmHg]
CVP
Systolic
14
6
10
2
10
2
Diastolic
6
-4
2
-4
2
-4
Mean
10
0
4
0
4
0
IBP 3
[mmHg]
LAP
Systolic
14
6
10
2
10
2
Diastolic
6
-4
2
-4
2
-4
Mean
10
0
4
0
4
0
IBP 4
[mmHg]
PAP
Systolic
35
10
60
24
60
24
Diastolic
16
0
4
-4
4
-4
Mean
20
0
26
12
26
12
IBP PR [bpm]
120
50
HIGH
160
75
HIGH
200
100
HIGH
EtCO2 [mmHg]
In
10
OFF
MIDDLE
10
OFF
MIDDLE
10
OFF
MIDDLE
Ex
40
10
MIDDLE
40
10
MIDDLE
40
10
MIDDLE
Multi
Gas
[vol%]
Gas1
In
12.0
OFF
HIGH
12.0
OFF
HIGH
12.0
OFF
HIGH
Ex
10.0
OFF
10.0
OFF
10.0
OFF
Gas2
In
12.0
OFF
12.0
OFF
12.0
OFF
Ex
10.0
OFF
10.0
OFF
10.0
OFF