VM15 User Manual Specification
239
NIBP (with SunTech Module) Optional
Method of Measurement
Oscillometric. Diastolic values correspond to Phase 5 Korotkoff sounds
Blood Pressure Range
Systolic
Adult: 40 ~ 260 mmHg
Pediatric: 40 ~ 160 mmHg
Neonate: 40 ~ 130 mmHg
Diastolic
Adult: 20 ~ 200 mmHg
Pediatric: 20 ~ 160 mmHg
Neonate: 20 ~ 100 mmHg
MAP (Mean)
Adult: 26 ~ 220 mmHg
Pediatric: 26 ~ 183 mmHg
Neonate: 26 ~ 110 mmHg
Pulse Rate Range
30 to 220 bpm (Beats Per Minute)
Pulse Rate Accuracy
± 2% or ± 3 bpm, whichever is greater
Cuff Deflate Rate
Deflation step size varies with heart rate, cuff pressure and cuff volume
Initial Inflation Pressure
Adult: 160 mmHg (default) / Variable from 120 to 280 mmHg
Pediatric: 140 mmHg (default) / Variable from 80 to 170 mmHg
Neonate: 90 mmHg (default) / Variable from 60 to 140 mmHg
Clinical Accuracy
Meets accuracy requirements of ANSI/AAMI
SP10:1992 and 2002
Pressure Transducer Accuracy
±3mmHg between 0 mmHg and 300 mmHg for
operating conditions between 0°C and 50°C
Recommended Frequency of Pressure Transducer Calibration
The Pressure Transducer calibration should be verified on a yearly interval
Operating Conditions
C to 50°C, 15% to 95% non-condensing humidity
Storage Conditions
-20°C to 65°C, 15% to 95% non-condensing humidity
Altitude
Measurement accuracy is not affected by altitude
Startup Initialization Period
7 seconds
Patient Safety
Internal operating software ensures that:
Maximum cuff inflation time is limited to 75 seconds
Duration of blood pressure reading is limited to
130 seconds (Adult mode)
120 seconds (Adult Motion Tolerant mode)
90 seconds (Pediatric mode)
75 seconds (Neonate mode)