VM15 User Manual Specification
239
NIBP (with SunTech Module) – Optional
⚫ Method of Measurement
◼ Oscillometric. Diastolic values correspond to Phase 5 Korotkoff sounds
⚫ Blood Pressure Range
◼ Systolic
◆ Adult: 40 ~ 260 mmHg
◆ Pediatric: 40 ~ 160 mmHg
◆ Neonate: 40 ~ 130 mmHg
◼ Diastolic
◆ Adult: 20 ~ 200 mmHg
◆ Pediatric: 20 ~ 160 mmHg
◆ Neonate: 20 ~ 100 mmHg
◼ MAP (Mean)
◆ Adult: 26 ~ 220 mmHg
◆ Pediatric: 26 ~ 183 mmHg
◆ Neonate: 26 ~ 110 mmHg
◼ Pulse Rate Range
◆ 30 to 220 bpm (Beats Per Minute)
◼ Pulse Rate Accuracy
◆ ± 2% or ± 3 bpm, whichever is greater
◼ Cuff Deflate Rate
◆ Deflation step size varies with heart rate, cuff pressure and cuff volume
◼ Initial Inflation Pressure
◆ Adult: 160 mmHg (default) / Variable from 120 to 280 mmHg
◆ Pediatric: 140 mmHg (default) / Variable from 80 to 170 mmHg
◆ Neonate: 90 mmHg (default) / Variable from 60 to 140 mmHg
◼ Clinical Accuracy
◆ Meets accuracy requirements of ANSI/AAMI
SP10:1992 and 2002
◼ Pressure Transducer Accuracy
◆ ±3mmHg between 0 mmHg and 300 mmHg for
operating conditions between 0°C and 50°C
◼ Recommended Frequency of Pressure Transducer Calibration
◆ The Pressure Transducer calibration should be verified on a yearly interval
◼ Operating Conditions
◆ 0°C to 50°C, 15% to 95% non-condensing humidity
◼ Storage Conditions
◆ -20°C to 65°C, 15% to 95% non-condensing humidity
◼ Altitude
◆ Measurement accuracy is not affected by altitude
◼ Startup Initialization Period
◆ 7 seconds
◼ Patient Safety
◆ Internal operating software ensures that:
Maximum cuff inflation time is limited to 75 seconds
Duration of blood pressure reading is limited to
130 seconds (Adult mode)
120 seconds (Adult Motion Tolerant mode)
90 seconds (Pediatric mode)
75 seconds (Neonate mode)