VM15 User Manual Specification
240
◆ Additional redundant safety circuitry oversees normal operation and will override
to abort a reading if:
cuff pressure exceeds 300 mmHg (Adult & Pediatric modes) or 150 mmHg
(Neonate mode) at any time
the cuff has been inflated for 180 seconds (Adult & Pediatric modes) or 90
seconds (Neonate mode)
The Module meets all relevant parts of the following
Safety Standards
IEC60601-1:1997
IEC/EN60601-2-30:1999/2000
AAMI SP10:1992/2002
EN1060-1:1996
EN1060-3:1997