VM15 User Manual Explanation of ICO (Option)
164
UNIT Function
Set mmHg or kPa as unit of blood pressure
ZERO Function
I. Do Zeroing function when measure for the first time.
II. Select IBP-ZERO for Zeroing function.
III. Ok is on the screen in case of successful Zeroing.
Fail is on the screen in case of failing Zeroing.
IBP ALARM Function
3 Kinds of Alarms: SYS, DIA, MEAN
Setting Range (SYS, DIA, MEAN, Alarm):
⚫ HIGH Range: -50 ~ 350mmHg, OFF
⚫ LOW Range: OFF, -50 ~ 350mmHg
Type Setting: HIGH, MEDIUM, LOW, OFF.
PR ALARM Function
3 kind of Alarm Steps: HIGH, LOW and TYPE
⚫ High alarm set: 0~300, Off
⚫ Low alarm set: Off, 0~300
Type alarm set: High, Medium, Low, Off