EasyManua.ls Logo

VOTEM VM15 - Explanation of ICO (Option)

VOTEM VM15
255 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
VM15 User Manual Explanation of ICO (Option)
165
24. Explanation of ICO (Option)
WARNING: Do not use any ICO cables or sensors other than those verified by
Edward company and supplied by VOTEM.
Heart (Left or Right Ventricular) for 1 minute to export this amount of blood "CO - Cardiac
output" is called.
CO units of L / min is.
As a time of ventricular contraction is the release of blood, the amount of SV (Stroke Volume)
is called,
The formula CO = SV x HR will be established.
Thermodilution measurement of the use of ICO, and is measured by calculating the following:
CO = V x (TB - TI) x (SI x CI) x 60 x CT x K
A (SB x CB) 1
WHERE:
CO = cardiac output
V = volume of injectate (ml)
A = area of thermodilution curve in square mm divided by paper speed (mm/sec)
K = calibration constant in mm/°C
TB, TI = temperature of blood (B) and injectate (I)
SB, SI = specific gravity of blood and injectate
CB, CI = specific heat of blood and injectate
(SI x CI ) = 1.08 when 5% dextrose is used
(SB x CB)
60 = 60 sec/min
CT = correction factor for injectate warming
Catheter's the Edwards Swan-Ganz Standard Thermodilution Pulmonary Artery Catheter Use.
To have ICO popup screen,
Select SCREEN smart button ( ) ICO