VM15 User Manual NIBP Monitoring
150
STAT Function
STAT is to repeat the measuring blood pressure at maximum times during the setting time.
Time set: 5 ~ 15 minute
NIBP STAT SCREEN appears when the use selects Screen-On. The measured values are
displayed ins serial order.
Select Screen - Off and NIBP STAT Screens is off.
PRESSURE Function
Set the MAX PRESSURE: 100~300 (Adult, Pediatric), 70~150 mmHg(Neonate)
Set the VENOUS STATSIS: 50~200 mmHg
UNIT Function
Change the unit if necessary : mmHg or kPa.
ALARM Function
ALARM has three kinds : SYS, DIA , MEAN , TYPE
NIBP measures the 3 different data. (SYS, DIA, MEAN). ALARM SET should be done for each
data.
High : 0~275, OFF
Low : OFF, 0~275.
Type : HIGH, MEDIUM, LOW, OFF.
TREND Function
The function of supplying the stored Trend Data on the screen.