VM15 User Manual NIBP Monitoring
149
Description of NIBP functions
NIBP Menu appears when the user selects NIBP Measurement Screen in Numeric area.
Press X and NIBP menu disappears. (Refer to the below Figure 20-4.)
Figure 21-5. DISPLAYS - NIBP Menu
MODE Function
Kinds of Setting: Manual, Auto, STAT, VENOUS STAT
I. Manual mode is for manual measurement.
II. Auto mode is for automatic measurement.
III. Refer to STAT Function for details.
IV. VENOUS STAT helps to find vein by controlling the flow of venous blood giving the
constant pressure to the patient who do now show vein in normal conditions.
NOTE
Control the flow of venous blood for 170 seconds (Adult, Pediatric), 90
seconds (neonate)
The user can set VENOUS STAT from 50 to 200mmHg
(every 5mmHg unit).
Default value is 160mmHg.
INTERVAL Function
In NIBP AUTO MODE, the user can set the time interval for next measurement.
The user can set 2, 3, 4, 5, 10, 15, 30 Minute, 1, 2, 4, 8, 12Hours.