VM15 User Manual Specification
237
SpO2 (with Masimo SET) – Optional
⚫ Measurement Range
◼ SpO2
◆ Display Range: 0 ~ 100 %
◆ Calibration Range: 70 ~ 100 %
◼ Pulse Rate
◆ Display Range: 25 ~ 240 bpm
◆ Calibration Range: 25 ~ 240 bpm
◼ Perfusion Index
◆ Display Range: 0.02 ~ 20.0 %
◆ Calibration Range: 0.10 ~ 20.0 %
⚫ Accuracy
◼ SpO2
◆ No Motion: ≤ 2.0 %
◆ Motion(RMS): ≤ 3.0 %
◼ Pulse Rate
◆ No Motion: ≤ 3.0 bpm
◆ Motion(RMS): ≤ 5.0 bpm
⚫ Resolution
◼ SpO2: ≤ 0.1 %
◼ Pulse Rate: ≤ 1 bpm
⚫ Time to Display
◼ SpO2: ≤8, ≤12 seconds
◼ Pulse Rate: ≤8, ≤12 seconds
◼ Perfusion Index: ≤8, ≤12 seconds
⚫ Asystole Detection Time
◼ SpO2: ≤8 seconds
◼ Pulse Rate: ≤8 seconds
◼ Perfusion Index: ≤8 seconds
⚫ Delay
◼ SpO2: ≤10 seconds
◼ Pulse Rate: ≤10 seconds
◼ Perfusion Index: ≤10 seconds
⚫ Response Time
◼ SpO2: ≤20 seconds
◼ Pulse Rate: ≤20 seconds
◼ Perfusion Index: ≤20 seconds
⚫ Setting
◼ Sensitivity: Maximum, Normal, APOD
◼ Averaging Time (seconds): 2~4, 4~6, 8, 10, 12, 14, 16 seconds
◼ Fastsat : ON/OFF
◼ Smart Tone: ON/OFF
◼ SIQ: ON/OFF
⚫ Alarm Delay (seconds): 0, 5, 10, 15 seconds
⚫ Rapid Desat Alarm (%): 5, 10, OFF