VM15 User Manual Masimo SpO2
Monitoring (Option)
142
PULSE SEARCH
MESSAGE
Device is searching for pulse.
If device fails to display within
30 seconds, disconnect and
reconnect sensor to patient. If
pulse search continues, move
sensor to better perfused site.
Specifications
SpO2, Pulse Oximetery (Masimo)
Table 20-7. SpO2, Pulse Oximetery (Masimo) Specification
Parameter
Functional SpO2
(%)
Pulse Rate (BPM)
Perfusion
Index (%)
Pleth
Variability
Index (%)
Display Range
0~ 100%
25 ~ 240BPM
0.02 ~ 20.0%
0~ 100%
Calibration Range
70~ 100%
25 ~ 240BPM
0.10 ~ 20.0%
-
Calibration
Standard
Invasive CO-
Oximeter
ECG and Patient
Simulator
Patient
Simulator
-
No Motion
Accuracy(RMS)
<2.0%
<3.0BPM
-
-
Motion
Accuracy(RMS)
<3.0%
<5.0BPM
-
-
Resolution
<0.1%
<1BPM
<0.01%
<1%
Time to Display
<8, <12 seconds
<8
5
, <12
6
seconds
<8
5
, <12
6
seconds
-
Asystole
Detection Time
<8 seconds
<8 seconds
<8 seconds
-
Delay
<10 seconds
<10 seconds
<10 seconds
-
Response Time
<20 seconds
<20 seconds
<20 seconds
-
Display Update
Frequency
>1 Hz
>1 Hz
>1 Hz
>1 Hz
Averaging Time
(seconds)
2-4, 4-6, 8, 10,
12, 14, 16
-
-
-
Alarm Range
High limit 30 ~
100%
SpO2 (Unit: 1%)
Low limit 20 ~
90%
SpO2 (Unit: 1%)
High limit 30 ~
240BPM (Unit:
1BPM)
Low limit 25 ~
230BPM (Unit:
1BPM)