VM15 User Manual Suntech NExplanation of ICO (Option)
155
MODE Function
Kinds of Setting: Manual, Auto, STAT, VENOUS STAT)
I. Manual mode is for manual measurement.
II. Auto mode is for automatic measurement.
III. Refer to STAT Function for details.
IV. VENOUS STAT helps to find vein by controlling the flow of venous blood giving the
constant pressure to the patient who do now show vein in normal conditions.
CAUTION: Do not apply to the normal patient. (It is only for abnormal cases.).
CAUTION: After VENOUS STAT Function, reset into Auto or Manual to measure blood
pressure.
Control the flow of venous blood for 170 seconds (Adult, Pediatric), 90
seconds (neonate)
The user can set VENOUS STAT from 50 to 200mmHg
(every 5mmHg unit).
Default value is 160mmHg.
INTERVAL Function
In NIBP AUTO MODE, the user can set the time interval for next measurement.
The user can set 2, 3, 4, 5, 10, 15, 30 Minute, 1, 2, 4, 8, 12Hours.
STAT Function
STAT is to repeat the measuring blood pressure at maximum times during the setting time.
Time set: 5 ~ 15 minute
NIBP STAT SCREEN appears when the use selects Screen-On. The measured values are
displayed ins serial order.
Select Screen - Off and NIBP STAT Screens is off.