VM15 User Manual Menu Default Set
209
31. Menu Default Set
If you change the patient category in the Default set menu, some settings are changed as
shown in the table below.
Table 31-1. Menu Default Set
Parameter
Item Name
Patient Category
Adult
Pediatric
Neonate
ECG
LEAD
1 CH
LEAD I
2 CH
LEAD II
3 CH
LEAD III
GAIN
10 mm/mV
SPEED
25 mm/sec
ST
J + 80ms
ARRHYMIA
OFF
PACEMAKER
OFF
FITER
Monitoring
TM
(STD1)
LEAD
LEAD II
GAIN
5 mm/mV
SPEED
12.5 mm/sec
SpO2
(VOTEM)
GAIN
x1
SPEED
25 mm/sec
AVERAGING
4
TM
(STD1)
GAIN
5 mm/mV
SPEED
12.5 mm/sec
SpO2
(MASIMO)
GAIN
x1
SPEED
25 mm/sec
SENSITIVITY
NORMAL
AVERAGING
2 ~ 4
FASTSAT
OFF
SMART TONE
OFF
SIQ
OFF
ALARM DELAY
5 sec
RAPID DESAT ALARM
10 %
TM
(STD1)
GAIN
x0.5
SPEED
12.5 mm/sec
RESP.
GAIN
x1
SPEED
25 mm/sec
APNEA
10 sec
TM
(STD1)
GAIN
x0.5
SPEED
12.5 mm/sec