VM15 User Manual Specification
242
IBP (Invasive Blood Pressure) - Optional
Measurement Range: -50 ~ +350 mmHg
Channel: 2 channels or 4 channels
Accuracy: ± 1 mmHg or ± 2%, which is greater.
Auto zero balancing range: ± 200 mmHg
Unit: mmHg, kPa
Alarm Range
IBP 1~4
[mmHg]
Display
High
Low
Systolic
-50 ~ +350, OFF
OFF, -50 ~ +350
Diastolic
Mean
Label
ABP (Arterial Blood Pressure)
Artery group throughout the whole body (systemic)
CVP (Central Venous Pressure)
Central vein / atrium group
LAP (Left Arterial Pressure)
Intracranial group
PAP (Pulmonary Arterial Pressure)
Pulmonary artery group
Catheter
Utah (Optional)
Abbot (Not provided)
Edward (Not provided)
Bd (Not provided)
Medex (Not provided)
Philips (Not provided)