VM15 User Manual Specification
247
Multi Gas - Optional
Mainstream (Multi Gas)
Operating Temperature
IRMA AX+: 5 to 50 °C (41 to 122 °F)
Operating Humidity
< 40 hPa H2O (non-condensing) (95 %RH at 30 °C)
Operating Atmospheric Pressure
5251200 hPa
(525 hPa corresponding to an altitude of 4,572 m / 15,000 feet)
Power Supply
IRMA AX+: 4.5-5.5 VDC, max 1.4 W
(power on surge @ 5 V less than 350 mA during 200 ms)
Breath Detection
Adaptive threshold, minimum 1 vol% change in CO2 concentration
Respiration Rate
0 ~ 150 bpm, ± 1 bpm
Warm-up Time
IRMA AX+: < 20 seconds (agent identification enabled and full accuracy)
Rise Time
CO2 ≤ 90 ms
N2O ≤ 300 ms
HAL, ISO, ENF, SEV, DES ≤ 300 ms
Total System Response Time
< 1 second
Accuracy - Standard Conditions
(valid for dry single gases at 22 ± 5 °C and 1013 ± 40 hPa)
CO2:
0 to 15 vol% ± (0.2 vol% + 2% of reading
N2O:
0 to 100 vol% ± (2 vol% + 2% of reading)
SEV:
0 to 10 vol% ± (0.15 vol% + 5% of reading)
DES:
0 to 22 vol% ± (0.15 vol% + 5% of reading)
HAL, ISO, ENF:
0 to 8 vol% ± (0.15 vol% + 5% of reading)
Accuracy - All Conditions
(valid for all specified environmental conditions)
CO2: ± (0.3 kPa + 4% of reading)
N2O: ± (2 kPa + 5% of reading)
Agents: ± (0.2 kPa + 10% of reading)