VM15 User Manual Printing (Option)
191
Placement of IRMA Probe
When connecting IRMA probe to an infant patient circuit it is important to avoid a direct
contact between the IRMA probe and the infant's body.
If, for whatever the reason, the IRMA probe is in direct contact with any parts of the infant's
body an insulation material shall be placed between the IRMA probe and the body.
Pre-use check
Prior to connecting the IRMA airway adapter to the breathing circuit, verify gas readings and
waveforms on the monitor before connecting the airway adapter to the patient circuit.
Perform the tightness check of the patient circuit with the IRMA probe snapped on the IRMA
airway adapter.