VM15 User Manual Setting up the VM15
64
Table 7-5. Battery Lifetime Table
Temperature
Lifetime
0 ~ 20
9 months
21 ~ 30
6 months
31 ~ 45
3 months
Note: VM15 has the internal type battery. Please contact to suppler, if battery life time is
lower than 30% of usual.
Note: Battery Specification
·Voltage: 10.89V 5200mAh (MAIN) : Standard(1ea), Option(1ea)
·Voltage: 10.89V 2600mAh (TM) : Standard(1ea)
·Battery Type: Lithium-ion
Note: Battery Changing Period
The lifetime of the battery gets shorter if it is not 100% recharged for use.
Table 7-6. Change Period of the Battery
Charging /
Discharge Rate
Number of Using
Times
100 %
1200 times
50 %
500 times
30 %
200 times
Note:
·If the power does not work, please check the following
·Check the connection status between power cord and socket.
·Check the AC ground.
·Check the main power source and voltage.
WARNING: Check the regulatory voltage before connecting the AC power source to the
patient monitor. The damage occurs when it is out of allowance range.
WARNING: Wave noise may happen when the patient monitor operates with AC power
source. Please make the ground.
WARNING: Connecting AC power source with wet hands shall bring the electronic shock and
critical damage to the patient monitor
WARNING: If the battery and power source is separated, all of stored data (Trend, Alarm
Setting value, Time, Date) can be reset. Please separate the battery only when keeping the
patient monitor for long time without use.
WARNING: Do not disassemble the main unit to take out the internal battery. It may cause
critical damage to the user and the patient monitor.