6-24
IM DLM6054-01EN
Examples
Trigger source
More than
Trigger
t
T1
t > T1
Trigger source
Less than
Trigger
t
T1
t < T1
Trigger source
Between
Trigger
t
T1
T1 < t < T2
T2
Trigger source
Trigger
t
T1
t > T1
Time out
Trigger source
Out of Range
Trigger
t
T1
T2
Trigger
t < T1 or t > T2
The figures below show how the trigger point is different in the “More than” and “Time Out”.
More than
Time = 400 ns
Trigger
500 ns
100 ns
CH1
Trigger
500 ns
100 ns
CH1
400 ns
Time Out
Time = 400 ns
6.7 Triggering on a Pulse Width Condition