6-32
IM DLM6054-01EN
Examples
CH1
CH2
H
H
CH1
CH2
L
H
L
L
t
T1
t > T1
Trigger
CH3
Mode: More than; Clock source: CH3,
State: CH1 = H, CH2 = L, other channels = X, AND
Condition: False
Result of comparison with
the state conditions
Not met
Not met
Met
Not met
Not met
Met
Results of comparisons
that have been sampled
using the clock
Result of comparison with
the state conditions
H
H
CH1
CH2
Mode: More than; Clock source: None
State: CH1 = H, CH2 = L, other channels = X, AND
Condition: False
Trigger
CH1
CH2
L
H
L
L
t
Time
t > Time
Not met
Not met
Met
6.9 Triggering on a State Width Condition