8.2 Displaying Accumulated Waveforms
Procedure
DISPLAY Accumulate Menu
Press DISPLAY and then the Accumulate soft key to display the following menu.
Turns accumulation on and off
Set the accumulation mode (Count, Time).
Set the gradation mode (Color, Inten).
Set the intensity and the accumulation count (accumulation time).
Explanation
Ordinarily, momentary waveform anomalies are difficult to recognize because the display is updated
whenever the trigger is activated. The accumulated waveform display allows you to observe
momentary anomalies by continuing to display each acquired data waveform for a set time.
Turning Accumulated Waveform Display On and Off (Accum)
ON: Accumulated waveforms are displayed.
OFF: Accumulated waveforms are not displayed.
Accumulation Mode (Mode)
Count: The DL6000/DLM6000 displays the current waveform and a set number of waveforms before
it.
Time: The DL6000/DLM6000 continues to display acquired waveforms for a set period of time.
A
waveform’s intensity decreases as time passes.
Accumulation Count
When the accumulation mode is Count, set the number of waveform accumulations to a number
from 0 to the number of history waveforms. Waveforms will accumulate infinitely if you choose 0.
The default value is the maximum number of savable history waveforms.
Accumulation Time
When the accumulation mode is Time, you can choose to accumulate waveforms infinitely, or for
a period of time between 100 ms and 100 s. Waveforms will accumulate infinitely if you choose
Infinite. The default value is 100 ms.
Gradation Mode (Grade)
Color: The DL6000/DLM6000 indicates 15 frequency levels using different colors. The color changes
from blue to green to yellow to red to white as the frequency increases.
Inten: The DL6000/DLM6000 indicates frequency using dif
ferent intensity levels. You can set the
number of intensity values to a number from 1 to 20. The default value is 8.